Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa 7534-02: (1 year) CAL VER /5520 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 31 NUMBER OF LINES: 132 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 0 - 50 degC 1.005 DISP Humidity: <70% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Turn on the UUT. 1.008 HEAD {CONTINUITY} 1.009 DISP Set the Function Switch to Ohms. 1.010 DISP Connect the 5520A to the UUT as follows: 1.010 DISP [32] 5520A AUX HI to UUT V/OHM 1.010 DISP [32] 5520A AUX LO to UUT COM 1.010 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.010 DISP [32] 5520A NORMAL LO to UUT COM 1.010 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 1.010 DISP the NORMAL leads to the SENSE leads AT the UUT. 1.011 5520 0.0Z S CW 1.012 MESS Press the SELECT button and check for an audible tone. 1.012 MESS 1.013 EVAL Does the Audible Tone check work correctly? 2.001 HEAD {RESISTANCE} 2.002 HEAD RESISTANCE: {400 Ohm Range} 2.003 5520 400 100.0Z 0.3% 0.05/ CW 3.001 HEAD RESISTANCE: {4 kOhm Range} 3.002 5520 4 1.000kZ 0.3% 0.025/ CW 4.001 HEAD RESISTANCE: {40 kOhm Range} 4.002 5520 40 10.00kZ 0.3% 0.025/ CW 5.001 HEAD RESISTANCE: {400 kOhm Range} 5.002 5520 400 100.0kZ 0.3% 0.025/ CW 6.001 HEAD RESISTANCE: {4 MOhm Range} 6.002 DISP Connect the 5520A to the UUT as follows: 6.002 DISP [32] 5520A NORMAL HI to UUT VOLT/OHM 6.002 DISP [32] 5520A NORMAL LO to UUT COM 6.003 5520 4 1.000MZ 0.5% 0.025/ 2W 7.001 HEAD RESISTANCE: {40 MOhm Range} 7.002 5520 40 10.00MZ 1% 0.025/ 2W 8.001 HEAD {FREQUENCY} 8.002 DISP Set the Function Switch to Hz. 8.003 5520 5000H 0.02% 1U 1V SI 2W 9.001 5520 50.00kH 0.02% 0.01U 1V SI 2W 10.001 HEAD {AC VOLTAGE} 10.002 HEAD AC VOLTAGE: {400mV Range} 10.003 5520 400 350.0mV 0.75% 0.2U 100H SI 2W 11.001 5520 400 350.0mV 1.5% 0.4U 100H SI 2W 12.001 HEAD AC VOLTAGE: {4V Range} 12.002 5520 4 3.500V 0.75% 0.002U 100H SI 2W 13.001 HEAD AC VOLTAGE: {40V Range} 13.002 5520 40 35.00V 0.75% 0.02U 100H SI 2W 14.001 5520 40 35.00V 1.5% 0.04U 1000H SI 2W 15.001 HEAD AC VOLTAGE: {400V Range} 15.002 5520 400 350.0V 0.75% 0.2U 100H SI 2W 16.001 HEAD AC VOLTAGE: {750V Range} 16.002 5520 750 700V 0.75% 2U 100H SI 2W 17.001 HEAD {DC VOLTAGE} 17.002 HEAD DC VOLTAGE: {400mV Range} 17.003 5520 400 350.0mV 0.1% 0.05/ 2W 18.001 HEAD DC VOLTAGE: {4V Range} 18.002 5520 4 3.500V 0.1% 0.05/ 2W 19.001 HEAD DC VOLTAGE: {40V Range} 19.002 5520 40 35.00V 0.1% 0.05/ 2W 20.001 HEAD DC VOLTAGE: {400V Range} 20.002 5520 400 350.0V 0.1% 0.05/ 2W 21.001 HEAD DC VOLTAGE: {1000V Range} 21.002 5520 1000 1000V 0.2% 0.1/ 2W 22.001 HEAD {DIODE TEST} 22.002 DISP Set the UUT Function Switch to Diode. 22.003 M5520 RNGLK 33V 22.004 5520 0.500V 0.3% 0.002U 2W 23.001 M5520 * 23.002 HEAD {LO DC CURRENT} 23.003 HEAD LO DC CURRENT: {40mA Range} 23.004 5520 40 10.00mA 1% 0.02U 2W 24.001 HEAD LO DCCURRENT: {400mA Range} 24.002 5520 400 100.0mA 1.5% 0.2U 2W 25.001 HEAD {LO AC CURRENT} 25.002 DISP Press the SELECT button to select AC Amps. 25.003 HEAD LO AC CURRENT: {40mA Range} 25.004 5520 40 10.00mA 2% 0.05U 100H SI 2W 26.001 HEAD LO AC CURRENT: {400mA Range} 26.002 5520 400 100.0mA 2% 0.5U 100H SI 2W 27.001 HEAD {HI AC CURRENT} 27.002 DISP Connect the 5520A to the UUT as follows: 27.002 DISP [32] 5520A AUX HI to UUT 10A 27.002 DISP [32] 5520A AUX LO to UUT COM 27.003 HEAD HI AC CURRENT: {10A Range} 27.004 5520 10 1.90A 2% 0.05U 100H SI 2W 28.001 HEAD {HI DC CURRENT} 28.002 DISP Press the SELECT button to select DC Amps. 28.003 HEAD HI DC CURRENT: {10A Range} 28.004 5520 10 1.90A 1.5% 0.02U 2W 29.001 HEAD {TEMPERATURE} 29.002 DISP Connect the 5520A TC Terminals to the UUT using the 29.002 DISP K-TYPE thermocouple adapter. 29.003 M5520 TCREF 21degC 29.004 5520 0degC 0.2% 1U _K TC 2W 30.001 5520 500degC 0.2% 1U _K TC 2W 31.001 END